| - Before starting, fill out the logbook | - Before starting, fill out the logbook | ||
| - Prepare your sample on the preparation table, mount it on a stub | - Prepare your sample on the preparation table, mount it on a stub | ||
| - | - Log out if necessary, login | ||
| - Click ' | - Click ' | ||
| - | - Unfreeze | + | - Make sure CCD (bottom right screen) is unfrozen. |
| - Mount the stub onto the stage, use the elephant to set a safe height | - Mount the stub onto the stage, use the elephant to set a safe height | ||
| - | - Check CCD when closing chamber | + | - Keep an eye on CCD when closing |
| - Click ' | - Click ' | ||
| - | - Wait until a low enough final pressure | + | - Wait until a 1e-5 mbar is reached |
| - Choose HV and spot | - Choose HV and spot | ||
| - | - Click ' | + | - Click ' |
| - Adjust contrast and brightness | - Adjust contrast and brightness | ||
| - Focus (on the highest point of the sample) at a magnification > 2000 times | - Focus (on the highest point of the sample) at a magnification > 2000 times | ||
| - Link Z to FWD | - Link Z to FWD | ||
| - | - Drive sample up to 10-15 mm, check CCD, be prepared to hit 'Esc' | + | - Drive sample up to 10-15 mm, check CCD, keep finger above 'ESC-key' |
| - Focus well at a magnification >2000 times | - Focus well at a magnification >2000 times | ||
| - Link Z to FWD again | - Link Z to FWD again | ||
| - Drive sample up to the working height, check CCD, be prepared to hit ' | - Drive sample up to the working height, check CCD, be prepared to hit ' | ||
| - Set eucentric if necessary (see below) | - Set eucentric if necessary (see below) | ||
| + | - LOCK Z-axis. Now you can switch to single-view. | ||
| - You are in search mode, go to the feature of interest | - You are in search mode, go to the feature of interest | ||
| - Increase magnification | - Increase magnification | ||
| - Switch to UHR mode (for a nonmagnetic sample) | - Switch to UHR mode (for a nonmagnetic sample) | ||
| - | - Focus, correct astigmatism | + | - Focus, correct astigmatism, lens alignment |
| - Collect data | - Collect data | ||
| - | - When ready, click ' | + | |
| + | - Go to QUAD-view, unlock Z | ||
| + | | ||
| - Remove your specimen stub | - Remove your specimen stub | ||
| - | - Call next user or pump down in ' | + | - Call next user or pump down (don't keep chamber vented for a significant amount of time) |
| - | - Log out and close the log entry in the logbook | + | |
| - At the end of the day, switch off the displays | - At the end of the day, switch off the displays | ||
| - Go to your PC and transfer the data | - Go to your PC and transfer the data | ||
| Line 114: | Line 115: | ||
| ^ Material ^ Date ^ Sample ID ^ Process parameters | ^ Material ^ Date ^ Sample ID ^ Process parameters | ||
| - | | Pt | 20050415 | + | | Pt | 20050415 |
| |Pt | 20050601 | |Pt | 20050601 | ||