Line 18: Line 18:
  
 If you use a tilt sample holder you can use silverpaint to mount the sample. Use this sample holder only if the use has been demonstrated to you since it is //very easy// to damage the final lens with it (repair cost: >10000 Euro)! If you use a tilt sample holder you can use silverpaint to mount the sample. Use this sample holder only if the use has been demonstrated to you since it is //very easy// to damage the final lens with it (repair cost: >10000 Euro)!
 +
  
  
Line 27: Line 28:
  
   * Loading the sample   * Loading the sample
-    - Mount the sample holder on the transfer stick.+    - Wear powder free gloves. 
 +    - Mount the sample holder on the transfer stick (Don't use glue or paste).
     - Place the transferstick in front of the loadlock with the excentric hole //down//     - Place the transferstick in front of the loadlock with the excentric hole //down//
     - Push the red button. The loadlock will be evacuated     - Push the red button. The loadlock will be evacuated
Line 38: Line 40:
     - //carefully// turn the valve a quarter turn anti-clockwise, let it slide in and turn a quarter turn clockwise.     - //carefully// turn the valve a quarter turn anti-clockwise, let it slide in and turn a quarter turn clockwise.
     - Push the red button to vent the loadlock //while holding the transferstick so that it doesn't fall on the floor// (if you think it is rediculous that I write this you should now admire the cracks in the metal-doped, X-ray blocking glass)     - Push the red button to vent the loadlock //while holding the transferstick so that it doesn't fall on the floor// (if you think it is rediculous that I write this you should now admire the cracks in the metal-doped, X-ray blocking glass)
-    - If the transfer did not go very smooth or the pressure went up, put vacuum grease (next to the transfer stick) onto the transfer stick and slide it back and forth (you are still wearing gloves, so your hands will not get dirty)+    - If the transfer did not go very smooth or the pressure went up (strangely this is indicated by having no illumination of the current meter and you can't turn on main power), put vacuum grease (next to the transfer stick) onto the transfer stick and slide it back and forth (you are still wearing gloves, so your hands will not get dirty)
     - Put the transfer stick in the holder     - Put the transfer stick in the holder
   * Activate the System   * Activate the System