lindberg_blue_tube_oven
Differences
This shows you the differences between two versions of the page.
| Both sides previous revisionPrevious revisionNext revision | Previous revision | ||
| lindberg_blue_tube_oven [2022/11/24 10:06] – scholma | lindberg_blue_tube_oven [2023/03/31 09:20] (current) – [Oxygen and Argong tube oven annealing] scholma | ||
|---|---|---|---|
| Line 1: | Line 1: | ||
| - | ====== Oxygen tube oven annealing ====== | + | ====== Oxygen |
| - | The tube oven has a temperature range of up to 1100 °C and can fit 26mm diameter tubes. Temperature regulation is done by a Lindberg/ | + | The tube oven has a temperature range of up to 1100 °C and can fit 26mm diameter tubes. Temperature regulation is done by a Lindberg/ |
| + | Oxygen and Argon can be connected separately or mixed. Setting the gas flows is done via a needle valve separately for O2 and Ar. | ||
| ====== Notes before usage ====== | ====== Notes before usage ====== | ||
| Line 20: | Line 22: | ||
| * Al2O3: 1700°C | * Al2O3: 1700°C | ||
| - | ====== Safety ====== | ||
| - | * high temperature | ||
| - | * pressurized gas cylinders | ||
| ==== High temperature safety ==== | ==== High temperature safety ==== | ||
| Line 39: | Line 38: | ||
| * check that your tube and sample boat have a higher operating temperature specified than your process temperature | * check that your tube and sample boat have a higher operating temperature specified than your process temperature | ||
| * load sample and connect tube adaptors | * load sample and connect tube adaptors | ||
| - | * double check reducer is closed | + | * double check pressure reducers on the cylinders are closed |
| - | * open oxygen | + | * open cylinder(s) you need (make sure there is enough oxygen for you proces) |
| - | * set 1 bar on secundary | + | * set 1 bar on secondary |
| - | * set desired flow rate with needle valve | + | * set desired flow rate with needle valve for both gasses separately. |
| * setup heater program and start (see below on how to use the temperature controller) | * setup heater program and start (see below on how to use the temperature controller) | ||
| Line 49: | Line 48: | ||
| * if you operated in single setpoint mode: turn off oven manually | * if you operated in single setpoint mode: turn off oven manually | ||
| * wait until oven temp is below 50°C | * wait until oven temp is below 50°C | ||
| - | * close cylinder, keep flow on until pressure on primary and secundary | + | * close cylinders with the angle key on the top of the cylinder, keep flow on until pressure on primary and secondary |
| - | * close reducer and needle | + | * close manual valves, |
| - | * take out sample and put aluminium foil on parts to keep them clean | + | * take out sample and put a tube back to keep everything |
| ==== Using the UP150 temperature controller ==== | ==== Using the UP150 temperature controller ==== | ||
| + | {{:: | ||
| + | |||
| + | A typical program for us will have three segments, a ramp-up, a dwell and ramp-down segment.\\ | ||
| + | In the ramp-up segment you specify the desired temperature and how fast you want to reach this temperature (do not heat to quickly).\\ | ||
| + | In the dwell segment you set the time how long you want to stay at the desired temperature in hours.\\ | ||
| + | In the ramp-down segment you specify the desired temperature and how fast you want to reach this temperature, | ||
| + | |||
| + | === changing ramp-up segment === | ||
| + | * hold set/ent button for 3 sec to go into the menu | ||
| + | * with set/ent button navigate to PrG | ||
| + | * set PrG to 1 with ▲ | ||
| + | * set/ent button to confirm and enter the program | ||
| + | * with set/ent button navigate to SP1 | ||
| + | * change desired temperature with ▼ and ▲, set/ent button to confirm | ||
| + | * change Tn1 with ▼ and ▲ (displayed in hours) | ||
| + | * hold set/ent button 3 sec to go back to main display | ||
| + | |||
| + | Start program: hold ▼ for 3sec | ||
| - | You can operate the temperature controller in two modes: | + | Stop program: hold ▲ for 3sec |
| - | * single setpoint mode: the controller will go to the desired temperature and maintain the temperature indefinitely, | + | |
| - | * programming mode: the user can enter a series of setpoint and time values, the controller will follow the sequence. | + | |
| - | {{lindberg-blue_up150_temperature_controller_manual.pdf|Manual UP150 controller}} | + | {{:lindberg-blue_up150_temperature_controller_manual.pdf|Manual UP150 controller}} |
lindberg_blue_tube_oven.1669284410.txt.gz · Last modified: 2022/11/24 10:06 by scholma