| ==== Nikon ECLIPSE LV150 optical microscope ==== | ==== Nikon ECLIPSE LV150 optical microscope ==== | ||
| - | + | Manufacturers Manual: {{ : | |
| - | The optical microscope in the cleanroom is a Nikon ECLIPSE LV150 optical microscope. | + | |
| - | + | ||
| - | The manufacturers manual can be found here: {{ : | + | |
| === Lamp === | === Lamp === | ||
| The microscope lamp is a (LV-HL50W) 12V50W halogen lamp with uniform light intensity distribution across the filament area. Filament-based lamps have a finite life-time, so **make sure to turn off the light** when you are done with the optical microscope. | The microscope lamp is a (LV-HL50W) 12V50W halogen lamp with uniform light intensity distribution across the filament area. Filament-based lamps have a finite life-time, so **make sure to turn off the light** when you are done with the optical microscope. | ||
| - | === Acetone | + | === Stage === |
| - | Since acetone can dissolve many plastics, do not spill any acetone on the stage. If you want to observe a lift-off process, you could use put a small petri dish in a larger petri dish, so that if the acetone spills from the smaller petri dish, it will not land on the stage. | + | Since acetone can dissolve many plastics, |
| - | === Stage === | ||
| It sometimes occurs that the X/Y-stage is stuck. If it happens, and the linear actuators are ' | It sometimes occurs that the X/Y-stage is stuck. If it happens, and the linear actuators are ' | ||
| - | If the stage does not move at all, it might be that the nut came loose from the spindle. This is usually the case when the nut is not tightened enough. | + | If the stage moves freely, the nut came loose from the spindle. This can happen |
| - | The cover panels can be removed to check if this is the case, and one could try and tighten the nut on the spindle. | + | (In the case that the Y-actuator is not working, it is easier to remove the stage from the mount. The nut cannot be seen from the top, so by removing the stage from the microscope, it is a bit easier to turn the stage upside down and inspect the nut-spindle situation. When removing the stage, be //extremely careful// as not to hit the objective lenses with the stage and potentially damaging them. It would be best to remove the nose of objective lenses first. How this is done, can be found in the manual linked above.) |
| - | (In the case that the Y-actuator is not working, it is easier to remove the stage from the mount. When doing this, be extremely careful as not to hit the objective lenses with the stage and potentially damaging them. It would be best to remove the nose of objective lenses first. How this is done, can be found in the manual linked above.) | + | |
| Sometimes, when the actuators are not operating properly, or failing every two or three days, the actuators just need a bit of grease. | Sometimes, when the actuators are not operating properly, or failing every two or three days, the actuators just need a bit of grease. | ||
| - | |||
| === Saving an image with annotations === | === Saving an image with annotations === | ||