oxides_system
Differences
This shows you the differences between two versions of the page.
| Next revision | Previous revision | ||
| oxides_system [2004/12/05 21:50] – (old revision restored) 127.0.0.1 | oxides_system [2013/02/27 09:09] (current) – boltje | ||
|---|---|---|---|
| Line 1: | Line 1: | ||
| + | ====== Description ====== | ||
| + | The oxides sputtering system is a epitaxial-growth high-pressure non-magnetron sputtering system with specially adapted sources that operate at pressures up to 5 mbar! The background pressure is typically <1e-6 mbar after a night of pumping and 1e-7 mbar is the minimum. The system is equipped with a 1200 C substrate heater that can rotate over the sources. Oxides are sputtered reactively, a 3-gas blending system for Ar, 0< | ||
| + | |||
| + | ====== Manual ====== | ||
| + | |||
| + | ====Safety==== | ||
| + | Safety issues for yourself and for the system: | ||
| + | * Before sputtering, make sure that **all power supply cables are connected** | ||
| + | * When **pumping down, switch off cooling water** for the sources and x-tal | ||
| + | * When starting, make sure the interlocks are wired correctly. **Do not plug 2 power supplies in the connector for the same cooling water loop!** | ||
| + | * After that, **check flow meters** and switch on cooling water for the sources and x-tal | ||
| + | |||
| + | |||
| + | [[oxide_YBCO]] | ||
| + | ==== Change Sample ==== | ||
| + | |||
| + | === Switch off the pumps === | ||
| + | - Check whether **all** gas inlet valves are closed | ||
| + | - Close the gate valve | ||
| + | - Switch off the pumps with the central pump switch | ||
| + | - Switch off the cooling water | ||
| + | |||
| + | === Open the chamber === | ||
| + | - Vent the chamber | ||
| + | - open N< | ||
| + | - turning the gas selector valve to N< | ||
| + | - Close both valves when the pressure is 8e2 Torr. | ||
| + | - Switch on the '' | ||
| + | - Raise the top of the system using the lift control unit and rotate the top to the left. | ||
| + | |||
| + | === Mount the substrate === | ||
| + | - **ALWAYS** wear gloves when touching substrates and system parts | ||
| + | - Remove the heater shield | ||
| + | - Remove samples, clean sideplates in HNO< | ||
| + | - Clean the heater surface | ||
| + | - with a razor | ||
| + | - 400 grit sandpaper | ||
| + | - 800 grit sandpaper | ||
| + | - Scotchbrite | ||
| + | - Clean with IPA. This step is crucial! | ||
| + | - Mount the sample with '' | ||
| + | - Cut away glue around the substrate | ||
| + | - Mount side plates, push them flush against the substrate using a cotton tip. | ||
| + | - Set the heater to 80 C for 1 hour using the Eurotherm control software | ||
| + | - Dry for >1 hour | ||
| + | - Mount the heater shield, blow possible dust particles away with N< | ||
| + | |||
| + | === Close the chamber === | ||
| + | - Rotate the top above the chamber and lock it in place | ||
| + | - Lower the top very carefully making sure that the teflon guides slide //along// the chamber (take into account safety rules for working with heavy objects) | ||
| + | - Switch off the lift main switch | ||
| + | |||
| + | === Pump down === | ||
| + | - Open the gate valve | ||
| + | - Switch on diaphragm pump | ||
| + | - Wait until the pressure is about 5 torr on the pirani or 8 mbar on the capacitance gauge | ||
| + | - Switch on cooling water | ||
| + | - Switch on turbo | ||
| + | - At 800 Hz: open main gas valve under the chamber | ||
| + | - Check that the ion gauge switches on | ||
| + | - Pump overnight | ||
| + | |||
| + | |||
| + | ==== Deposition ==== | ||
| + | == Procedure start== | ||
| + | - Pressure <1E-6 mbar | ||
| + | - Position heater between 2 sources, opposite to Target | ||
| + | - Open **N< | ||
| + | - Open the O< | ||
| + | - Adjust Turbo speed to regulate pressure | ||
| + | - Open the Target shutter, keep other shutters closed (unless you make a multilayer of course) | ||
| + | |||
| + | === RF === | ||
| + | - check cooling water | ||
| + | - Switch on power supply, power regulation | ||
| + | - Switch on matching power supply | ||
| + | - Check for no reflection when turning power up | ||
| + | - Let everything stabilize | ||
| + | |||
| + | === DC === | ||
| + | - Switch on the target power supply, current regulation | ||
| + | - Set the setpoint at 8 ma, display actual current and voltage | ||
| + | - when pressure is right, switch power supply to on. The plasma should ignite. Get help if it doesn' | ||
| + | - Let the pressure increase until the voltage gets below 350-380 V (the lower the better!) | ||
| + | - The voltage keeps decreasing, keep it constant at 350-380 V by increasing current (it is critical that you follow this rate given by the O< | ||
| + | - Keep increasing current, set current to 350 mA (never get above 390 V!) | ||
| + | |||
| + | == Presputtering == | ||
| + | - You will see that the conditioning of the target takes quite a while, check that while conditioning the voltage goes down to about the previous value in the logbook. | ||
| + | - Presputter for >1 hour | ||
| + | - Program the heater in the meanwhile. | ||
| + | |||
| + | == Sputtering == | ||
| + | - Run temperature the recipe | ||
| + | - When temperature is reached, turn sample above target. | ||
| + | - Start a timer. | ||
| + | - Time the epitaxial growth process | ||
| + | |||
| + | == After Sputtering == | ||
| + | - Turn off power supply | ||
| + | - Cool down sample | ||
| + | |||
| + | === End === | ||
| + | - Close O< | ||
| + | - Let pump run for 20 minutes | ||
| + | - Close N< | ||
| + | - Wait until the sample temperature is <50 C | ||
| + | - Check whether **all** gas valves are closed | ||
| + | |||
| + | ==== Recipes ==== | ||
| + | |||
| + | ====== Deposition rates ====== | ||
| + | |||
| + | ^ Material ^ Date ^ Sample ID ^ Process parameters | ||
| + | | LCMO | 20060918 | ||
| + | | Pt | 20120711 | ||
| + | | LAO | 20121210 | ||
| + | |||
| + | |||
| + | ====== Tech stuff ====== | ||
| + | * Targets: 2 inch OD, 0.125 inch thick | ||
| + | * Backingcup: 53mm OD, 6.3mm thick | ||
| + | |||
| + | |||