 ===== Plasma Cleaner manual ==== ===== Plasma Cleaner manual ====
 {{ :plasma_cleaner_pdc-32g-2_usermanual_2022.pdf |Plasma Cleaner manual}} {{ :plasma_cleaner_pdc-32g-2_usermanual_2022.pdf |Plasma Cleaner manual}}
 +
 ==== DO's and DONT's ==== ==== DO's and DONT's ====
  
-Only use bare substrate in the plasma cleaner!+Only use bare substrate in the plasma cleaner, all exceptions should be discussed.
   * Si   * Si
   * SiO2   * SiO2
   * glass/quartz   * glass/quartz
-all exceptions should be discussed 
- 
-==== Checking system status and before starting ==== 
  
  
 ==== MANUAL ==== ==== MANUAL ====
  
-  * stop pump +  * Load samples on the sample tray 
-  * turn valve to the right to vent +  * Start the vacuum pump while holding the door, make sure black valve is pointing down 
-    * vent slowly to avoid samples flying around in the chamber +  * Pump down to <100mTorr 
-  * load samples +  * Set black valve to the needle valve 
-  * close vent valve (pointing down) +    * if needle valve is open the pressure should go to previous setpoint (check the logbook)
-  * start pump, hodor +
-  * pump down to <100mTorr +
-  * set valve to needle valve +
-    * should be going to previous setpoint (check logbook)+
     * or change the setpoint     * or change the setpoint
-  * wait for pressure to stabilize+  * Wait for pressure to stabilize
   * set RF (low, mid or high)   * set RF (low, mid or high)
-  * when plasma ignites start timer +  * When plasma ignites start your timer 
-  * turn off plasma when timer finishes +  * Turn off RF when timer finishes 
-  * close valve (down position+  * Close black valve (valve pointing down) 
-  * pump down to <100mTorr +  * Pump down to <100mTorr to make sure etch products are removed from the chamber 
-  * stop pump +  * Stop the vacuum pump 
-  * vent +  * Turn black valve to the right to vent 
-  close vent valve and pump down +    * vent slowly to avoid samples flying around in the chamber 
- +  * Close black valve (valve pointing down 
- +  * close the door 
-==== Finishing up and leaving the system ==== +  * write in the logbook: namedate, sample, process conditions, any remarks
- +
- +
- +
-==== Common problemstips and tricks ==== +
- +
- +
 ==== Recipes ==== ==== Recipes ====
-typical process:\\+Typical cleaning process:\\
 500-900mTorr\\ 500-900mTorr\\
 1-3 minutes\\ 1-3 minutes\\
 High RF High RF
  
- 
-==== Available accesoiries and tools ==== 
  
 === Maintenance === === Maintenance ===
  
 +Tip seal change of vacuum pump\\
 +Cleaning sample tray and chamber\\
  
  
plasma_cleaner_pdf-32g-2.1675347579.txt.gz · Last modified: 2023/02/02 14:19 by scholma