stm:blog
Differences
This shows you the differences between two versions of the page.
| Both sides previous revisionPrevious revisionNext revision | Previous revision | ||
| stm:blog [2006/09/12 09:55] – galli | stm:blog [2007/01/03 12:15] (current) – galli | ||
|---|---|---|---|
| Line 1: | Line 1: | ||
| + | ====== STM Blog ====== | ||
| + | (This is a kind of logbook page. You can report daily progress, if you find it relevant for the STM users) | ||
| + | |||
| + | **03-02-2006**\\ | ||
| + | The RHK SPM 1000 Rev.8E control electronics has been modified in order to reduce the output XYZ, XYZ-offsets voltages from +/- 215V to +/- 140V. This has been done in order to protect scanning tubes thinner than 1mm from being depoled by too high voltages. All the details will be soon uploaded under the RHK electronics wiki page. | ||
| + | |||
| + | |||
| + | **06-02-2006**\\ | ||
| + | The first RT, on the table, imaging of the topography of a " | ||
| + | |||
| + | {{stm: | ||
| + | |||
| + | **07-02-2006**\\ | ||
| + | Now the STM is hanging vertical. Other images of the same film. | ||
| + | |||
| + | {{stm: | ||
| + | |||
| + | **08-02-2006 and 09-02-2006**\\ | ||
| + | STM is cold. Images at 200K (08020634.png) and at 100K (09020602.png). | ||
| + | |||
| + | {{stm: | ||
| + | |||
| + | **10-02-2006**\\ | ||
| + | Tested small biases (3mV instead of 300mV used for the above pictures). Tested " | ||
| + | |||
| + | {{stm: | ||
| + | These images are at small bias (3mV) instead of 300mV used for the previous ones. Used " | ||
| + | |||
| + | {{stm: | ||
| + | These images are at small bias (3mV) and the " | ||
| + | |||
| + | **13-02-2006**\\ | ||
| + | Are these terraces or artefacts (not enough Z-position resolution)? | ||
| + | |||
| + | {{stm: | ||
| + | |||
| + | I-V spectroscopy, | ||
| + | Note the large V-offset (-400mV): check this out! | ||
| + | |||
| + | {{stm: | ||
| + | |||
| + | **15-02-2006**\\ | ||
| + | |||
| + | STM out. Tested magnet with Cryogenics Limited power supply.\\ | ||
| + | Set Field=2.0 Tesla, Driven mode, OK.\\ | ||
| + | Set Persisten Mode, OK.\\ | ||
| + | Set Field=0.8 Tesla, Driven, OK.\\ | ||
| + | Set Persistent Mode, OK.\\ | ||
| + | Set ZERO. | ||
| + | |||
| + | **17-02-2006**\\ | ||
| + | |||
| + | Au/Mica (Au-100nm, sputtered by Ruud Hendrikx in the ATC, about 1 month ago).\\ | ||
| + | Images on the table reveal a not so flat sample. Nevertheless on the top of the plateaus the surface is very flat. Unfortunately some 415Hz distorted some of the higher res. images. When the STM was inserted in the cryostat the 415Hz made scanning impossible. Internal mechanical vibration is suspected. Took the STM out. | ||
| + | |||
| + | {{stm: | ||
| + | |||
| + | **07-03-2006**\\ | ||
| + | |||
| + | Cooled down STM with a Au/Mica sample and using a 0.25mm Titanium spring (" | ||
| + | |||
| + | {{stm: | ||
| + | |||
| + | **09-03-2006**\\ | ||
| + | |||
| + | Cooled down STM, after taking it out to air and changing sample (HOPG, now). The Z motor works again, down to 5K (did not go lower). Here after IV on HOPG, 200K (used " | ||
| + | |||
| + | {{stm: | ||
| + | |||
| + | **03-04-2006**\\ | ||
| + | |||
| + | 18:15, 7.88 Tesla: The Great White Plume...\\ | ||
| + | |||
| + | {{stm: | ||
| + | |||
| + | **12-04-2006**\\ | ||
| + | |||
| + | I-V curves on some places of LSMO(62nm)/ | ||
| + | |||
| + | {{stm: | ||
| + | |||
| + | **25-04-2006**\\ | ||
| + | |||
| + | Cryostat warmed up for Easter. Pumped vacuum spaces last week, precooled with LN2 yesterday.\\ | ||
| + | Today, when expelling LN2 from He-tank the following issue arised: nitrogen gas (in large quantities) was ending up in the helium recovery line (gas meter rotating fast). After a quick debugging I relised the N2 gas was entering the helium recovery line from the λ-plate pumping port. Either there' | ||
| + | |||
| + | **22-06-2006**\\ | ||
| + | |||
| + | Line scan of the superconducting gap of a BSCCO crystal (optimally doped) at 4K. | ||
| + | |||
| + | {{stm: | ||
| + | |||
| + | |||
| + | **25-08-2006**\\ | ||
| + | |||
| + | Nice results on LCMO/ | ||
| + | |||
| + | {{stm: | ||
| + | |||
| + | **28-08-2006**\\ | ||
| + | |||
| + | Problems with the magnet power supply model SMS120C. Contacted Cryogenics Ltd. Here is the email about the issue: | ||
| + | |||
| + | {{stm: | ||
| + | |||
| + | Using now SMS80C (max. 80 Amp.) controller from AMC group while waiting for repair of SMS120C.\\ | ||
| + | |||
| + | **10/ | ||
| + | |||
| + | Measuring LCMO on NGO (strain free, ~ 50nm). Measurements on insulating NGO substrate are possible. | ||
| + | Discoveries: | ||
| + | |||
| + | * Sharp tips are critical requirement for stable and long (2 weeks) measurements. | ||
| + | * If tips are cut lousily (very easy to do that), minitips form, and tunneling from different minitips occurs. This is of course bad for imaging (see tip mirroring features here under) and spectroscopy (noisy and unreliable). | ||
| + | * Tested etched W-tip. Good, but not long lasting (due to oxidation). | ||
| + | * Discovered the " | ||
| + | * Mitra et al. T-dependence of G(0) confirmed, although contamination (condensation of water et al.) makes measurements occasionally less sensitive than in a UHV environment (large tunneling gaps are then required). | ||
| + | |||
| + | |||
| + | {{stm: | ||
| + | |||
| + | **05-12-2006**\\ | ||
| + | |||
| + | Problems with the He probe of the green cryostat. Probe broken. Contact Cryogenic | ||
| + | |||
| + | {{stm: | ||