| Line 183: | Line 183: | ||
| - A low-ohmic short between the bias-signal and the bias-shield (later found, in the SSMC connector down at the STM, insert side). | - A low-ohmic short between the bias-signal and the bias-shield (later found, in the SSMC connector down at the STM, insert side). | ||
| - No energy resolution for Nb-gaps and smaller. Probably related to previous point (to be able to continue the experiments the bias-shield was interrupted just before it entered the cryostat, maybe introducing too much bias noise). | - No energy resolution for Nb-gaps and smaller. Probably related to previous point (to be able to continue the experiments the bias-shield was interrupted just before it entered the cryostat, maybe introducing too much bias noise). | ||
| - | - Problem with Au surface preparation (later related to mostly | + | - Problem with Au surface preparation (later |
| + | Federica, Marius Trouwborst | ||
| + | **20 April 2011 - 18 August 2011**\\ | ||
| + | * Repaired the wobble-stick (Chris Pen) | ||
| + | * Repaired the short (Marius, Federica). The short was inside one of the 2 SSMC connectors at the end of the refrigerator, | ||
| + | * Did a number of experiments at RT primarily to be sort out the Au <110> surface treatment for the missing-row reconstruction. Raphael Mueller, bachelor student from Germany, sorted out the procedure successfully.\\ | ||
| + | Really nice STM images at room temperature in HV (~2x10-8 mbar) of Au< | ||
| + | {{stm: | ||
| + | {{stm: | ||
| + | Really nice STM images at room temperature in HV (~2×10-8 mbar) of Au< | ||
| + | |||
| + | Initially, after the 1st cycle of 2 x sputter/ | ||
| + | |||
| + | {{stm: | ||
| + | |||
| + | But after more sputter/ | ||
| + | |||
| + | {{stm: | ||
| + | {{stm: | ||
| + | {{stm: | ||
| + | |||
| + | Cooled down STM. Cryogenic run starts. Operators: Federica Galli, Marius Trouwborst (postdoc) and Raphael Mueller (bachelor student).\\ | ||
| + | Coarse approach is troublesome: | ||
| + | |||
| + | Missing row reconstruction on Au< | ||
| + | |||
| + | {{stm: | ||
| + | {{stm: | ||
| + | {{stm: | ||
| + | {{stm: | ||
| + | |||
| + | |||
| + | **2012 - 2013**\\ | ||
| + | |||
| + | System was mostly warm for maintenance and several " | ||
| + | |||
| + | * Original Oxford lead-screw was replaced with a new high quality electrochemically polished lead-screw from Dutch company XXX which makes screws for ASML machines | ||
| + | * Original bellow-rotary drive on top of cryostat, to actuate lead-screw rotation, was replaced with UHV Deigns (UK company) " | ||
| + | * A complete new manipulation system based on 3 concentrical Magi Drives was developed by C. Pen (FMD) and F. Galli to be able to rotate the STM insert in the desired position when down in tip/sample change position. This to achieve 2 things: a) position the STM in perfect alignment with the wobble stick to facilitate tip/sample exchange and b) to be able to rotate the STM insert towards the evaporator (K-cell). | ||
| + | * replaced old twisted stainless steel rods of the insert with new ones | ||
| + | * installed a new Si3N4 triangular slider (machined and polished by Ceratec) in the STM. This caused 2 consecutive problems during cool-down possible because: different thermal expansion coefficient between Ti-body and silicon nitride slider, geometrical mis-match of the slider (it was delivered too big), not very good surface finish. We then decided to go back to a new (undamaged) Ti/TiNC slider, having trust in the fact that with the new manipulator tip and sample changes would be less distructive for the coating. | ||
| + | * Replaced all springs in the STM with CuBe springs instead of phosphor bronze. | ||
| + | * Installed a "hot plug" K-cell in the main chamber | ||
| + | * Installed Cu-powder RF filters from Cryolinks in the cold right above the STM for bias and tunneling current lines. | ||
| + | * Installed π-filters at room-temperature outside vacuum, on all electrical lines entering the system. | ||
| + | |||
| + | All pumps and o-rings maintenance.\\ | ||
| + | |||
| + | In July 2013, cool-down to 300mK was sucessfull. No measurement was performed, only test cool-down. STM was kept closed. | ||
| + | |||
| + | **September 2013 - December 2013**\\ | ||
| + | |||
| + | Cryogenic run. Operators: F.Galli, Christian Wagner (Juelich), Sumit Tewari (Leiden, JvR), Sasha Vrbica (Leiden, JvR).\\ | ||
| + | |||
| + | |||
| + | SUMIT SUMMERIZE FALL 2013 CRYOGENIC RUN WITH SOME TXT AND NICE PICS!\\ | ||
| + | |||
| + | At the end of the cool-down a first attempt was made to evaporate gold atoms on the cold sample in the STM in the " | ||
| + | |||
| + | So the decision was taken to warm up over the Xmas break to exclusively change the K-cell position (needed venting of main chamber). | ||
| + | |||
| + | **January 2014 - December 2014**\\ | ||
| + | |||
| + | Changed the evaporator position from bottom tilted to horizontal (slight grazing angle evaporation geometry).\\ | ||
| + | |||
| + | Cryogenic run (February 2014). Operators: F.Galli, Christian Wagner (Juelich), Sumit Tewari (Leiden, JvR), Sasha Vrbica (Leiden, JvR).\\ | ||
| + | |||
| + | SUMIT AND SASHA SUMMERIZE 2014 CRYOGENIC RUN WITH SOME TXT AND NICE PICS!\\ | ||
| + | |||
| + | **January 2015 - June 2015**\\ | ||
| + | |||
| + | System warmed up for maintenance and to fix a few shorts which developed over time. | ||
| + | System develops a leak in the OVC (1st bottom flange) during 1st cool-down in April because of heavy corrosion at the flanges. Repaired. | ||
| + | |||
| + | Several improvements in the electrical diagram, see attached PDF (Sumit!) | ||
| + | |||
| + | **July 2015 - Mid-August 2015**\\ | ||
| + | |||
| + | Cool-down. Coarse approach works well with external drive (HV amplifier and Sumit' | ||
| + | Repaired by Sumit and Chris. | ||
| + | |||
| + | **September 2015 - **\\ | ||