 ===== Coarse approach working principle ===== ===== Coarse approach working principle =====
  
-COARSE APPROACH: This STM employs a compact design where the Z-coarse and Z-fine motion of the tip is done by the same set of (shear) piezos. The same HV amplifier is used for Z-coarse and Z-fine. This means that the available voltage range has/can to be reserved for the Z-coarse and Z-fine motion, as  depicted in the figure here under:\\+COARSE APPROACH: This STM employs a compact design where the Z-coarse (a slip-stick motor, 6 piezo stacks driven at the same time with the same waveform) and Z-fine motion of the tip is done by the same set of (shear) piezos. The same HV amplifier is used for Z-coarse and Z-fine. This means that the available voltage range has/can to be reserved for the Z-coarse and Z-fine motion, as  depicted in the figure here under:\\
  
 {{stm:approach.png?400}}\\ {{stm:approach.png?400}}\\
Line 19: Line 21:
   - APPROACH: Waveform: Parabola, Asymmetry: 0%   - APPROACH: Waveform: Parabola, Asymmetry: 0%
   - RETRACT: Waveform: Sawtooth, Asymmetry: 100%   - RETRACT: Waveform: Sawtooth, Asymmetry: 100%
 +Typical cycle time is 200-250 μs.
  
  