| Line 113: | Line 113: | ||
| * set other process parameters with the GUI | * set other process parameters with the GUI | ||
| * load temperature program from itools to the controller and run the program | * load temperature program from itools to the controller and run the program | ||
| + | * hold the run/hold button for a few seconds to stop the controller, if the controller is in hold state the program will not load. | ||
| Line 146: | Line 147: | ||
| * put blind flanges or aluminium foil on the open parts of the system and tube | * put blind flanges or aluminium foil on the open parts of the system and tube | ||
| + | |||
| + | ===== Common problems ===== | ||
| + | |||
| + | === PLC not connected === | ||
| + | The extra network card in the PC sometimes loses connection. Open the PC and unplug and plug back the network card. Restart the computer. Open the PLC program and try to connect to the PLC. LOGO!SoftComfort will recognize that the IPadres of the PLC is not in the same subnet mask and will ask to add this; acknowledge. Now you should have connection again. | ||