uhv_system
Differences
This shows you the differences between two versions of the page.
| Both sides previous revisionPrevious revisionNext revision | Previous revision | ||
| uhv_system [2017/10/09 12:22] – [Venting the loadlock and loading the substrates:] scholma | uhv_system [2025/10/29 16:07] (current) – doorn | ||
|---|---|---|---|
| Line 1: | Line 1: | ||
| ====== Description ====== | ====== Description ====== | ||
| + | The UHV system is a magnetron sputtering system with a background pressure around 10e-10 mbar and a deposition pressure of 1e-3 to 1e-2 mbar. It is equipped with a loadlock that holds a sample garage with 7 sample spaces. One sample space can hold a 15x15mm wafer. The substrate is transferred to and from the chamber using two linear manipulators and a wobble stick. Most targets in the chamber are at an angle with respect to the sample, and the sample can be tilted towards 3 out of 4 magnetron sources. The system can be used with both DC and RF power supplies (DC being standard). The sputtering gas is Ar; the loadlock is also connected to N< | ||
| - | The UHV system is a magnetron sputtering system with a background pressure in the | + | The UHV system is maintained by the nanolab |
| - | 10e-10 mbar range and a deposition pressure of 1e-3 to 1e-2 mbar. It is equipped | + | |
| - | with a loadlock that holds 6 15x15 mm substrates. The substrate is transferred to | + | |
| - | and from the chamber | + | |
| - | be tilted towards 3 from the 4 2" magnetron sources. The system | + | |
| - | RF power supplies. Sputtering gas is Ar, the | + | |
| - | loadlock is also connected | + | |
| - | ====== | + | ======The Vacuum System====== |
| + | To keep the chamber conditions UHV, it is never vented to atmospheric pressure. Instead, samples are first pumped down to 5e-8 mbar inside a small loadlock. Samples are transferred from loadlock to chamber and back while the chamber is at an Ar overpressure with respect to the loadlock. This ensures contaminants from the loadlock cannot reach the chamber. | ||
| - | ===== Before | + | The pressures in loadlock and chamber are read out via ion gauges on the black panel below the system. IG1 reads out the chamber, while IG2 reads out the loadlock (always double check with stickers). When the light is on, the gauge is on and it is also being read out. **Ion gauges can explode when operated at high pressures, so you should never vent or let in any gas with the ion gauge on.** A third gauge is located on one of the chamber sides and is used to gauge the chamber pressure when Ar flows into the chamber. It automatically switches between an ion gauge and a pirani gauge, so that it works at sputter pressures. |
| - | Check the logbook to see if you are interfering with someone else. Fill in your name, date and time in the logbook before you start. Also mark in the logbook in which positions you are going to put the substrates (1 to 6). | + | The system also has intermediate vacuum gauges behind |
| - | ===== Venting the loadlock and loading the substrates: | + | ====== Manual ====== |
| - | + | ||
| - | - Wear //powder free// gloves! | + | |
| - | - Clean the substrates with demi-water (if necessary), then acetone and finally IPA | + | |
| - | - Glue the substrates with a small amount of silver paste on the sample holders. Wait 10 to 20 mins. for the silver paste to dry | + | |
| - | - Turn off the UHV chamber and loadloack pressure gauges (IG1 and IG2 respectively, | + | |
| - | - Close the argon line valve (left green valve) | + | |
| - | - Check if the gate valve between the UHV chamber and the loadlock is closed before venting the loadlock | + | |
| - | - In order to vent the loadlock, close the valve (large black knob) of the rotary pump then switch off the turbo pump of the loadlock. Wait 1 minute and then open the N< | + | |
| - | - When the turbo has stopped, completely vent and open the loadlock | + | |
| - | - Put the sample holders in the loadlock with the long side towards you. Always wear gloves! You can use long metal tweezers to hold the lip of the sample holder and put in the sample. | + | |
| - | - Close the loadlock | + | |
| - | - Close the N< | + | |
| + | ===== Before you start ===== | ||
| + | First check the logbook to see if you are interfering with someone else. Users write down their names upon loading their samples. So fill in your name and the date before starting your loading procedure! | ||
| + | ===== Venting the loadlock and loading the substrates: ===== | ||
| + | - **Never touch anything inside the machine without wearing clean, powder free gloves.** | ||
| + | - Sampleholders are kept on the workbench in the sputterlab, inside the petri dishes. Always clean your sampleholder with acetone and IPA before starting. | ||
| + | - Glue your sample onto the sampleholder with a small amount of silver paste. Let it dry (approximately 10 minutes). You should be able to hold your precious sample upside down! | ||
| + | - Turn off the loadlock ion gauge (IG2). Monitor IG1 while venting. If it starts increasing, you are probably making a mistake. | ||
| + | - Check if the gate valve between loadlock and chamber is fully closed. Clockwise = close. | ||
| + | - Turn off the loadlock turbo on the turbo controller panel. | ||
| + | - Close the valve of the rotary pump. Clockwise = close. This closes off the rotary pump. | ||
| + | - Let the turbo spin down for 1 minute. Then very slowly and carefully let in a bit of N< | ||
| + | - When the turbo has stopped, vent the system completely with N< | ||
| + | - Put your sampleholder inside the garage with the long side towards you. Always wear gloves, and use long metal tweezers. Take care to maintain N< | ||
| + | - Close the loadlock. The door needs to be in perfect alignment, so tilt it a little if uncertain. You should see absolutely no gap between the loadlock and the door. As you close the door, also close the N< | ||
| ===== Pumping down the loadlock: ===== | ===== Pumping down the loadlock: ===== | ||
| - | + | | |
| - | | + | - The pressure should start decreasing almost immediately. If it does not decrease, close the rotary pump valve and check if the door is properly shut. |
| - | - When the pressure on B (the intermediate pressure gauge of the loadlock) reaches 1e-1 mBar, switch on the loadlock turbo pump | + | - When the pressure on A (the loadlock |
| - | - When the pressure on B reaches | + | - When the turbo reaches |
| - | - Open the argon line valve (left green valve) | + | - Pump until base pressure (<5e-8 mbar) in the loadlock |
| - | - Pump until the base pressure | + | |
| ===== Sputtering: ===== | ===== Sputtering: ===== | ||
| **Pre-sputtering** | **Pre-sputtering** | ||
| - | - Fill in the log book. | + | - Fill in logbook |
| - | - Turn on the cooling water | + | - Turn on cooling water. There are two valves at the back of the instrument (blue labels on the left). |
| - | - Take care that the sources | + | - Take care that the sources are connected to the DC switch box. They could be disconnected, |
| - | | + | - Turn off all IGs on the black panel. |
| - | - Switch the IG1 pressure gauge off before you let in the Argon (Ar) gas! Otherwise you will destroy it | + | - Open the shutter that you are going to use. Make sure all other shutters |
| - | - Switch off the sublimation pump | + | - **Fully close the NV.** The needle |
| - | - Open the shutter | + | - Close the valve conductance |
| - | - Close the loadlock argon line valve (left) and open the chamber argon line valve (right) | + | - Turn on the argon-measuring gauge (Pfeiffer controller on the rack). Give it a minute |
| - | - Open the green Ar valve on the gas-panel | + | - Slowly open the argon line on the black panel. |
| - | - Close partially | + | - Gradually open the NV and let in ~2e-3 mbar of argon pressure. It takes some “human PID” to get this right. N.B.: even when the NV is fully closed, there is still a substantial flow through |
| - | - Adjust | + | - Now use the valve-conductance controller to regulate the pressure for your sputtering process. |
| - | - Put the setpoint | + | - Select your target |
| - | - Make sure that also the shutter | + | - Power on the MDX-1K DC power source. **Never operate |
| - | - Close the window shutters | + | - Set the setpoint |
| - | - Start presputtering | + | - Close all window shutters! Also the one for the light. |
| - | - When you finished with presputtering | + | - Ignite the plasma by pressing output on. **Never switch the output |
| + | - When finished with presputtering | ||
| + | - Turn off the power supply. If you want to sputter multiple materials, pre-sputter every material, taking extra care that you are opening/closing the right shutters. | ||
| + | |||
| **Loading** | **Loading** | ||
| - | - Open the gate valve between the loadlock and the chamber (you will see the pressure decreasing a little, because | + | - Open the gate valve. The argon will also flow into the loadlock turbo now. This will decrease |
| - | | + | - Make room for the garage by moving the sample holder far away from the center |
| - | - Open the shutter | + | - Put the garage |
| - | - Use the manipulator system of the loadlock to insert the substrates | + | - Slowly and carefully take out your sample from the garage with the manipulator |
| - | - Use the manipulator | + | - Now drive the sample holder into a comfortable |
| - | - Use the wobble stick to position | + | - Use the manipulator to “put down” your sample. Be careful that the beak does not smack your sample out of the holder. |
| - | - Use the manipulator | + | - Now place your sample in the middle of the chamber. The correct |
| - | - Close the gate valve between | + | |
| - | - Move the substrate holder until the substrate | + | |
| **Sputtering** | **Sputtering** | ||
| - | - Tilt the subtrate holder if you use the front or back target (+45 or -45 degrees). | + | - If necessary, tilt your sample to the right sputtering angle. |
| - | - Close the Window Shutters. | + | - Check the pressure. If necessary, adjust |
| + | - Check the shutters. | ||
| - Check that you have the right Ar-Pressure, | - Check that you have the right Ar-Pressure, | ||
| - | - Turn on the crystal monitor (XTC), press zero, press stop, press start, again press start as at the same time you turn on the power supply | + | - Close the window shutters |
| - | - When you have sputtered | + | - Update |
| - | - For multiple layers, repeat the required steps but remember to close the shutters | + | - Press output on and time your sputtering process |
| + | - Monitor | ||
| + | - When finished, press output off. | ||
| + | - Turn off the power supply. If you want to sputter a different material, start again with pre-sputtering. | ||
| + | - Close all used shutters. | ||
| **Finishing** | **Finishing** | ||
| - | - Close the shutters of all the targets and open the Window Shutters | + | - Drive back your sample to a place where you can easily grab it with the manipulator |
| - | - Open again the gate valve between the loadlock | + | - Grab your sample with the manipulator |
| - | - Use the wobble stick to put the sample back on the loadlock holder (be careful not to scratch the film against the loadlock holder) | + | - Now open the gate valve. Ar pressure should decrease again. Drive the garage into a comfortable position |
| - | - Return the manipulator in the loadlock. Close the gate valve between | + | - If you need to sputter multiple samples, exchange sample and sputter again. |
| - | - If you are done, switch the XTC off | + | - Drive back the garage and close the gate valve fully. Ar pressure should return to the level you set before. |
| - | - Open the valve of the turbo pump 100%. | + | - Close the Ar valve on the gas panel fully. |
| - | - Close the Ar valve on the gas-panel fully | + | - First pump out whatever gas remains, by opening |
| - | - Open the needle | + | - Open the NV fully. Once the dual-gauge goes below 5e-6 mbar, the chamber is empty and you can now turn it off. |
| - | | + | - Turn on IG1 again, and check that the chamber is returning to UHV conditions. |
| - | - Close Window Shutters | + | - Switch off cooling water. |
| - | - Switch off the rack of the power supplies | + | |
| - | - Switch off the cooling water and switch on the sublimation pump | + | |
| - Vent the loadlock as described above in order to remove your samples | - Vent the loadlock as described above in order to remove your samples | ||
| - | - Pump down the loadlock | + | - Always pump down the loadlock after use (as described above). |
| - | | + | |
| + | ===== Bake-out: ===== | ||
| + | To achieve a high-quality vacuum after the chamber has been vented, a bake-out is required. A good bake-out should take at least 48 hours to properly heat the whole system. | ||
| + | |||
| + | | ||
| + | * Disconnect the HV cables from the targets | ||
| + | * Remove the Ar gauge | ||
| + | * Cover the viewports | ||
| + | * Place all the panels to fully cover the whole machine | ||
| + | |||
| + | The blue cables for the vacuum gauges do not need to be covered. To start the bake-out, flip the breaker switch up and turn the black knob to the left. This permanently turns on the heaters (will go between the two setpoints, should be 80 and 100 C). Turning it to the right will use a timer but this was found to be iffy. After the bake-out make sure to reattach all cables and the Ar gauge. Please be aware that there is no interlock preventing you from turning on the heaters without the cooling water. The cooling water must be on during the bake-out! | ||
| ====== Deposition rates ====== | ====== Deposition rates ====== | ||
| + | |||
| + | RATES AFTER THE BIG MAINTENANCE OF 2025 | ||
| + | |||
| + | All rates measured by Jibbe Reuver with tapping mode-AFM in June 2025. | ||
| + | |||
| + | ^ Material ^ Process parameters ^ Thickness ^ Rate (nm/min) ^ Rate (nm/ | ||
| + | | Co | 4.0e-3 mbar, 100 mA, 10 min | 31.2 nm | 3.12 nm/min | 0.052 nm/s | | ||
| + | | Cu | 4.0e-3 mbar, 65 mA, 10 min | 136.86 nm | 13.69 nm/min | 0.228 nm/s | | ||
| + | | Nb | 4.0e-3 mbar, 200 mA, 20 min, 45 deg | 61.58 nm | 3.08 nm/min | 0.0513 nm/s | | ||
| + | | Pt | 4.0e-3 mbar, 150 mA, 10 min, 45 deg | 122.9 nm | 12.29 nm/min | 0.205 nm/s | | ||
| + | | Pt | 4.0e-3 mbar, 100 mA, 10 min | 111.37 nm | 11.14 nm/min | 0.1856 nm/s | | ||
| + | | Pt | 4.0e-3 mbar, 120 mA, 10 min | 138.53 nm | 13.85 nm/min | 0.2309 nm/s | | ||
| + | | Pt | 4.0e-3 mbar, 100 mA, 10 min, 45 deg | 87.8 nm | 8.78 nm/ | ||
| Definition: '' | Definition: '' | ||
| - | RECENT | + | **Small note on rate calibrations: |
| + | |||
| + | RATES BEFORE THE BIG MAINTENANCE OF 2025 | ||
| ^ Material | ^ Material | ||
uhv_system.1507551726.txt.gz · Last modified: 2017/10/09 12:22 by scholma