| === How to clean the wedge === | === How to clean the wedge === | ||
| - | 10min in the ultrasonic bath in ISO-Propanol FIXME | + | If the wedge is clogged, you might want to clean it. |
| + | Find a small petridish and fill it with a bit of isopropanol. Hold the petridish underneath the wedge without spilling, and use the Test-switch to turn ultrasounds on. This way, we basically ultrasonicate the wedge as if it were in a ultrasonicator. Sometimes, 10 seconds is enough to clean the wedge. If it is still clogged, you might want to remove it. Please contact a technician before you do that. | ||
| - | ===== Available accesoiries and tools ===== | ||
| - | |||