Plasma Cleaner manual

DO's and DONT's

Only use bare substrate in the plasma cleaner, all exceptions should be discussed.

  • Si
  • SiO2
  • glass/quartz

MANUAL

  • Stop the vacuum pump
  • Turn valve to the right to vent
    • vent slowly to avoid samples flying around in the chamber
  • Load samples
  • Close vent valve (valve pointing down)
  • Start the vacuum pump, hodor
  • Pump down to <100mTorr
  • Set valve to the needle valve
    • if needle is open pressure should go to previous setpoint (check the logbook)
    • or change the setpoint
  • Wait for pressure to stabilize
  • set RF (low, mid or high)
  • When plasma ignites start your timer
  • Turn off RF when timer finishes
  • Close valve (valve pointing down)
  • Pump down to <100mTorr
  • Stop the vacuum pump
  • Vent (valve pointing right)
  • Close vent valve (valve pointing down)
  • Pump down

Common problems, tips and tricks

Recipes

Ttypical cleaning process:
500-900mTorr
1-3 minutes
High RF

Maintenance

Tip seal change of vacuum pump