Plasma Cleaner manual
DO's and DONT's
Only use bare substrate in the plasma cleaner, all exceptions should be discussed.
- Si
- SiO2
- glass/quartz
MANUAL
- Stop the vacuum pump
- Turn valve to the right to vent
- vent slowly to avoid samples flying around in the chamber
- Load samples
- Close vent valve (valve pointing down)
- Start the vacuum pump, hodor
- Pump down to <100mTorr
- Set valve to the needle valve
- if needle is open pressure should go to previous setpoint (check the logbook)
- or change the setpoint
- Wait for pressure to stabilize
- set RF (low, mid or high)
- When plasma ignites start your timer
- Turn off RF when timer finishes
- Close valve (valve pointing down)
- Pump down to <100mTorr
- Stop the vacuum pump
- Vent (valve pointing right)
- Close vent valve (valve pointing down)
- Pump down
Common problems, tips and tricks
Recipes
Typical cleaning process:
500-900mTorr
1-3 minutes
High RF
Maintenance
Tip seal change of vacuum pump