stm:logbook_uhv_stm
This is an old revision of the document!
UHV He3 STM Logbook
Log here the cooldowns: Name_setup/Date-start/He-level start/He-level end/Sample/T-range/Magnetic field range/Date-end/Notes. Under “Notes” report whether He has been refilled during the cooldown. End date is defined when the sample/insert is taken out to air.
| Name STM/setup | Name | Date start (dd/mm/yy) | He-start (mm) | He-end (mm) | Sample | T-range (K) | UHV pressure (mbar) | H-range (T) | Date-end (dd/mm/yy) | Notes |
|---|---|---|---|---|---|---|---|---|---|---|
| S-STM | federica | 06-02-07 | 0 | 0 | several | 300 | 6.6 10-8 | none | 18-02-07 | initial testings |
| S-STM | federica | 19-02-07 | 0 | 0 | MoGe/Au(50nm) old cap layer | 300 | 6.6 10-8 | none | 22-02-07 | further testing of the scanner. |
| S-STM | federica | 26-02-07 | 0 | 0 | MoGe/Au(50nm) 230207 | 300 | 6.6 10-8 | none | 16-03-07 | further testing. Problems with signal stability. No scanning possible. Change tip. Succeded, but when mounting the sample back the sample-holder holder comes loose! UHV have to opened again! |
| S-STM | federica | april-07 | 0 | 0 | n.a. | 300 | A.P. | none | end may-07 | repair of sample and tip holders. “Upgrade” of the Z-motor spacers into a frame that provides mechanical stop mechanism to avoid spring deformation during tip exchange. New coarse approach “policy”, see blog. |
| S-STM | federica | -06-07 | 0 | 0 | several | 300 | pumping down | none | closed the system. Pumping down |
stm/logbook_uhv_stm.1181650941.txt.gz · Last modified: 2008/12/11 16:22 (external edit)