Measurement Procedure
Starting point: new sample and new tip are loaded in the STM. Tip-sample distance is “large” (i.e. few millimeters). STM is in lowest position, inside the centering pin and 1K-shield-window is open.
Coarse and fine approach procedure with STM in lower position:
- Position and focus USB camera such that you can see clearly sample and tip (start software “ ” on Labview/T-control computer)
- Femto preamp: check back sidde of RHK, switch to BNC. Start Labview VI: STM Conductance. Set correct gain.
- RHK premp: check back side of RHK, switch towards D9.
- Prepare RHK controller: XYZ offsets in the middle (5.0). Set initially: time constant ~0.2, gain ~0.3. Set appropriate bias and current set points.