afm_room_626_tapping_mode
Differences
This shows you the differences between two versions of the page.
| Both sides previous revisionPrevious revisionNext revision | Previous revision | ||
| afm_room_626_tapping_mode [2007/05/23 14:33] – hendrikx | afm_room_626_tapping_mode [2007/07/04 12:29] (current) – hendrikx | ||
|---|---|---|---|
| Line 1: | Line 1: | ||
| + | AFM room 626 tapping mode (version May 2007) | ||
| + | |||
| + | |||
| + | ===== Principle ===== | ||
| + | |||
| + | |||
| + | The most commonly used non-contact or partial contact technique, | ||
| + | The advantages of IC mode are (1) the brief contact time between the tip and the surface, which | ||
| + | effectively removes the stick-and-slip behavior; and (2) the average tip-surface distance remains in the non-contact regime, which reduces the surface deformation when there is brief contact. The disadvantage of this mode is that the images are acquired in the attractive regime, which is not as sensitive as the repulsive region used during contact mode imaging. In other words, the intrinsic resolution of IC mode is not as high as the contact mode for hard and non adhesive | ||
| + | surfaces. | ||
| + | |||
| + | |||
| + | ===== Hardware connections ===== | ||
| + | |||
| + | |||
| + | For tapping mode or IC mode the RHK SPM100 is extended with Stanford | ||
| + | |||
| + | On the front side of the SPM 100 the polarity switch should be changed from ” +” to “ –“. | ||
| + | |||
| + | ===== Cantilever mounting ===== | ||
| + | |||
| + | The cantilever should be mounted as described in the contact mode manual. The cantilever holder should be suited for tapping mode. | ||
| + | Before beginning, verify that the mode switch on the MultiMode SPM’s base is switched to AFM & LFM , not to TM! If the switch is toggled to STM, the laser will be turned off. | ||
| + | |||
| + | ===== Align the laser ===== | ||
| + | |||
| + | a. Find the edge of the substrate (X-axis adjustment knob).\\ | ||
| + | (Note: Ensure that you have reached the front edge of the substrate, which is different to the tip of the cantilever. You may need to follow along the edge of the substrate to reach the correct position.) \\ | ||
| + | b. Find the base of the cantilever (Y-axis adjustment knob). \\ | ||
| + | c. Move to the tip of the cantilever (X-axis adjustment knob). \\ | ||
| + | |||
| + | |||
| + | |||
| + | ===== Modulation Frequency ===== | ||
| + | |||
| + | First set the driving amplitude to 0.5-1.0 Volts. | ||
| + | The most straightforward method to determine the natural resonance frequency of a cantilever is to monitor the amplitude of the cantilever vibration with an oscilloscope while scanning the driving frequency (the maximum frequency the SR830 lockin amplifier can deal with is 100 kHz). When the cantilever is excited with a frequency near its resonance, the amplitude will start to increase. A typical cantilever, with k = 2.2 N/m typically exhibits a resonance at 82 kHz (see cantilever box). The frequency spectra provide a good guide for intermittent contact conditions. The modulation frequency (f) should very near the peak of the spectrum which is the natural frequency of the cantilever, a bit lower so that the amplitude is like 90% of the maximum . The amplitude should be sufficiently high (especially under ambient conditions) in order to overcome the capillary forces and pull the cantilever off the surface after brief contact (~1nm). | ||
| + | |||
| + | |||
| + | ===== Imaging ===== | ||
| + | |||
| + | Before approaching the tip: | ||
| + | |||
| + | 1. Adjust the offset in such a way that the offset is about half the value of the amplitude. | ||
| + | |||
| + | 2. On the scan head, adjust the PSD in such a way that both in both directions the signal is about 0, so no offset. | ||
| + | |||
| + | 3. Set the time constant of the lockin as fast as possible to achieve a stable output. Then the dominant time constant in the system will be the feedback loop bandwidth and not the lockin which could cause delays in responding. | ||
| + | |||
| + | In the measurement a relatively high time constant on the RHK box (upto 10.0) might be needed. However, | ||
| + | in the approaching procedure the time constant should be a low value, about 1.0. Otherwise the tip might crash. | ||
| + | |||
| + | |||
| + | |||
| + | |||
| + | |||
| + | |||
| + | ===== Typical settings ===== | ||
| + | |||
| + | |||
| + | software: line time 1 sec 500x500 nm,aux feedback enable, time constant 10 sec, output 0 | ||
| + | |||
| + | spm100: | ||
| + | |||
| + | lockin: | ||
| + | |||
| + | ===== Configuration file ===== | ||
| + | |||
| + | * AFM tapping mode \\ | ||
| + | | ||