User Tools

Site Tools


cressington_sputter_coater_208hr

Differences

This shows you the differences between two versions of the page.

Link to this comparison view

Both sides previous revisionPrevious revision
Next revision
Previous revision
cressington_sputter_coater_208hr [2015/08/07 15:14] steeniscressington_sputter_coater_208hr [2015/08/10 08:04] (current) steenis
Line 6: Line 6:
   - Lift the top plate and remove the glass cylinder. Take care that the top-plate does not fall forward. The target can be damaged by striking the sample table.   - Lift the top plate and remove the glass cylinder. Take care that the top-plate does not fall forward. The target can be damaged by striking the sample table.
   - Place your sample, mounted on a stub, on the stage.   - Place your sample, mounted on a stub, on the stage.
-  - Unlock the tilt lever of the stage +  - Unlock the tilt lever of the stage. 
-  - Set tilt at desired angle and lock the tilt lever again +  - Set tilt at desired angle and lock the tilt lever again. 
-  - If a thickness controlled process will be used change the angle of the thickness crystal so that it faces towards the sputter target +  - If a thickness controlled process will be used change the angle of the thickness crystal so that it faces towards the sputter target. 
-  - Check that the stage isn't hindered by the wire of the thickness crystal by letting the stage rotate at the desired speed+  - Check that the stage isn't hindered by the wire of the thickness crystal by letting the stage rotate at the desired speed.
   - Replace glass cylinder and gently lower the top plate. Check that the top plate sits correctly on the glass cylinder.   - Replace glass cylinder and gently lower the top plate. Check that the top plate sits correctly on the glass cylinder.
-  - Turn on the system and allow the system to pump to better than 0,01 mbar and then wait for another 5 minutes+  - Turn on the system and allow the system to pump to better than 0,01 mbar and then wait for another 5 minutes.
  
 +===== Automatic Controlled Process =====
 +==== Pre-Sputtering ====
 +   - Select Auto with the auto/manual button.
 +   - Select timer.
 +   - Set pre-sputtering time at 10 seconds by holding PAUSE/TEST and pressing the timer SET buttons.
 +   - Set pre-sputter current to desired value by holding the SET mA button and pressing the timer SET buttons. Available currents 20, 40, 60, and 80 mA.
 +   - Make sure the shutter is closed
 +   - Press CYCLE/STOP to start pre-sputering process. The system will proceed through its standard cycle.
 +
 +==== Sputtering ====
 +   - Select timer for a timed process or MTM for a thickness controlled process.
 +        - Timed process, Change sputtering time by holding PAUSE/TEST and pressing the timer SET buttons. Time can be set between 5 and 300 seconds.
 +        - Thickness controlled process
 +             - Turn on the thickness monitor (MTM 20)
 +             - Set the correct density according to the target material by pressing the DENSITY button once and find the correct setting by scrolling with the scroll buttons.
 +             - Set the correct tooling factor according to stage tilt by pressing the TOOLING button once and find the correct setting by scrolling with the scroll buttons.
 +             - Set desired coating thickness by holding the TERMINATOR button for one second. The value for the thickness will show up and can be adjusted with the scroll buttons. 
 +   - Set sputter current to desired value by holding the SET mA button and pressing the timer SET buttons. Available currents 20, 40, 60, and 80 mA.
 +   - Set roatating speed of stage at desired value.
 +   - Open shutter
 +   - Press CYCLE/STOP to start the sputtering process. The system will proceed through its standard cycle.
 +  - After sputtering note the final thickness for a thickness controlled process.
 +  - Close shutter.
 +
 +==== When Done ====
 +   - Switch off the system to automatically vent the chamber, which will take approximately 2 minutes.
 +   - Lift the top plate and remove the glass cylinder.
 +   - Set tilt of the stage to the horizontal position.
 +   - Remove your sample.
 +   - Clean inside of the glass cylinder with a tissue and ethanol to keep the glass clear.
 +   - Replace glass cylinder and gently lower the top plate. Leave the system vented.
 +
 +===== Manual process for Cr-sputtering =====
 +   - Select manual with the auto/manual button.
 +   - Select timer for a timed process or MTM for a thickness controlled process.
 +        - Timed process, Change sputtering time by holding PAUSE/TEST and pressing the timer SET buttons. Time can be set between 5 and 300 seconds.
 +        - Thickness controlled process
 +             - Turn on the thickness monitor (MTM 20)
 +             - Set the correct density according to the target material by pressing the DENSITY button once and find the correct setting by scrolling with the scroll buttons.
 +             - Set the correct tooling factor according to stage tilt by pressing the TOOLING button once and find the correct setting by scrolling with the scroll buttons.
 +             - Set desired coating thickness by holding the TERMINATOR button for one second. The value for the thickness will show up and can be adjusted with the scroll buttons.
 +   - Set sputter current to 80 mA by holding the SET mA button and pressing the timer SET buttons.
 +   - Set rotating speed of stage at desired value.
 +   - Open argon flush valve for 10 seconds, then close.
 +   - Pump to 0,001 mbar
 +   - Open argon flush valve for 3 seconds, then close
 +   - Open leak valve.
 +   - When pressure reaches better than 0,03 mbar set MTM-20 to ZERO for a thickness controlled process
 +   - Press START/STOP to start sputtering
 +   - Initially the discharge will be pink of colour. It will turn blue when the surface layer of chromium oxide is removed and chromium begins toe sputter.
 +   - When plasma turns blue open the shutter
 +   - After sputtering note the final thickness for a thickness controlled process.
 +   - Close shutter.
  
 ====== Available Targets ====== ====== Available Targets ======
-^  Target Material  ^  Atomic Number  ^  Density in g/cm3  ^+^  Target Material  ^  Atomic Number  ^  Density in g/cm³  ^
 |  Pt/Pd (80/20)  |  78/46  |  19,52  | |  Pt/Pd (80/20)  |  78/46  |  19,52  |
 |  Cr  |  24  |  7,19  | |  Cr  |  24  |  7,19  |
cressington_sputter_coater_208hr.1438960465.txt.gz · Last modified: 2015/08/07 15:14 by steenis