cressington_sputter_coater_208hr [2015/08/10 07:44] steeniscressington_sputter_coater_208hr [2015/08/10 08:04] (current) steenis
Line 44: Line 44:
    - Clean inside of the glass cylinder with a tissue and ethanol to keep the glass clear.    - Clean inside of the glass cylinder with a tissue and ethanol to keep the glass clear.
    - Replace glass cylinder and gently lower the top plate. Leave the system vented.    - Replace glass cylinder and gently lower the top plate. Leave the system vented.
 +
 +===== Manual process for Cr-sputtering =====
 +   - Select manual with the auto/manual button.
 +   - Select timer for a timed process or MTM for a thickness controlled process.
 +        - Timed process, Change sputtering time by holding PAUSE/TEST and pressing the timer SET buttons. Time can be set between 5 and 300 seconds.
 +        - Thickness controlled process
 +             - Turn on the thickness monitor (MTM 20)
 +             - Set the correct density according to the target material by pressing the DENSITY button once and find the correct setting by scrolling with the scroll buttons.
 +             - Set the correct tooling factor according to stage tilt by pressing the TOOLING button once and find the correct setting by scrolling with the scroll buttons.
 +             - Set desired coating thickness by holding the TERMINATOR button for one second. The value for the thickness will show up and can be adjusted with the scroll buttons.
 +   - Set sputter current to 80 mA by holding the SET mA button and pressing the timer SET buttons.
 +   - Set rotating speed of stage at desired value.
 +   - Open argon flush valve for 10 seconds, then close.
 +   - Pump to 0,001 mbar
 +   - Open argon flush valve for 3 seconds, then close
 +   - Open leak valve.
 +   - When pressure reaches better than 0,03 mbar set MTM-20 to ZERO for a thickness controlled process
 +   - Press START/STOP to start sputtering
 +   - Initially the discharge will be pink of colour. It will turn blue when the surface layer of chromium oxide is removed and chromium begins toe sputter.
 +   - When plasma turns blue open the shutter
 +   - After sputtering note the final thickness for a thickness controlled process.
 +   - Close shutter.
  
 ====== Available Targets ====== ====== Available Targets ======
-^  Target Material  ^  Atomic Number  ^  Density in g/cm3  ^+^  Target Material  ^  Atomic Number  ^  Density in g/cm³  ^
 |  Pt/Pd (80/20)  |  78/46  |  19,52  | |  Pt/Pd (80/20)  |  78/46  |  19,52  |
 |  Cr  |  24  |  7,19  | |  Cr  |  24  |  7,19  |
cressington_sputter_coater_208hr.1439192678.txt.gz · Last modified: 2015/08/10 07:44 by steenis