| Line 9: | Line 9: | ||
| ===== Vacuum control and turning on the e-beam ===== | ===== Vacuum control and turning on the e-beam ===== | ||
| - | Usually VegaTC is running continuously on the left screen. If it is not or if you have to reboot the computer, open it by clicking the shortcut on the desktop. You will see this window: | + | ==== Venting and pumping ==== |
| - | {{:: | + | Usually VegaTC is running continuously on the left screen. If it is not (or after a reboot), open it by clicking the shortcut on the desktop. You will see this window: |
| + | |||
| + | {{:: | ||
| Note that the beam in the bottom right corner will be turned off when you start a new patterning session and after loading a sample. | Note that the beam in the bottom right corner will be turned off when you start a new patterning session and after loading a sample. | ||
| Line 17: | Line 19: | ||
| In the centre you have the SEM image window, below that the measurement toolbar. If this toolbar isn't visible, you can make it show by **INSERT COMMAND HERE**. On the right side you have the control elements for the SEM, beam and vacuum. **Before venting the chamber**, note down the Column pressure in the log book, shown under '' | In the centre you have the SEM image window, below that the measurement toolbar. If this toolbar isn't visible, you can make it show by **INSERT COMMAND HERE**. On the right side you have the control elements for the SEM, beam and vacuum. **Before venting the chamber**, note down the Column pressure in the log book, shown under '' | ||
| - | After loading the sample, close the door and click '' | + | After loading the sample, close the door and click '' |
| - | Once in a while, VegaTC shows a message about auto-heating the filament after you click on '' | + | ==== Turning on the e-beam ==== |
| + | |||
| + | Once the pressure is below 2.5e-4mbar you can turn on the electron beam. If you have a little more time to spare (about 10 minutes more), turning on HV at lower pressures prolongs the filament life time. Turn on the gun by clicking on '' | ||
| + | |||
| + | Once in a while, VegaTC shows a message about auto-heating the filament after you click on '' | ||
| + | |||
| + | It is recommended to allow the gun to heat up for at least 30 minutes prior to patterning. If you notice large fluctuations in beam current (more than ±0.01nA) you might want to allow for even more time to heat up. Typically, 60 to 90 minutes does the trick **<CHECK WITH LUC OR MARCEL> | ||
| ===== e_line ===== | ===== e_line ===== | ||
| + | e_line is the software you use to do all the patterning and moving around the stage. Double-click on the e-line logo on the desktop to start e_line. | ||
| + | |||
| + | {{: | ||
| + | |||
| + | |||
| + | ==== Crashes and startup problems ==== | ||
| + | e_line can be sensitive at times. If e_line is already open from another user, close it and wait for a couple of minutes (5 to 10 minutes). The software needs this time to fully shut down in the background. If you try to re-open it too quick it might crash and mandate a reboot. | ||
| + | |||
| + | If e_line crashes and it doesn' | ||
| + | |||
| + | ===== Defining the origin ===== | ||
| + | Define the origin of the global coordinate system on your sample. Typically this is the lower left corner of your sample. | ||
| + | |||
| + | |||
ebpg_graphic_manual.1739104324.txt.gz · Last modified: 2025/02/09 12:32 by fabian