jeol_820_sem
Differences
This shows you the differences between two versions of the page.
| Both sides previous revisionPrevious revisionNext revision | Previous revision | ||
| jeol_820_sem [2008/04/16 15:16] – Explained abreviation WD, corrected typfout, corrected italic end verstoep | jeol_820_sem [2008/05/29 15:59] (current) – verstoep | ||
|---|---|---|---|
| Line 18: | Line 18: | ||
| If you use a tilt sample holder you can use silverpaint to mount the sample. Use this sample holder only if the use has been demonstrated to you since it is //very easy// to damage the final lens with it (repair cost: >10000 Euro)! | If you use a tilt sample holder you can use silverpaint to mount the sample. Use this sample holder only if the use has been demonstrated to you since it is //very easy// to damage the final lens with it (repair cost: >10000 Euro)! | ||
| + | |||
| + | |||
| Line 25: | Line 27: | ||
| {{images: | {{images: | ||
| - | - Mount the sample holder on the transfer stick. | + | |
| - | - Place the transferstick in front of the loadlock with the excentric hole //down// | + | - Wear powder free gloves. |
| - | - Push the red button. The loadlock will be evacuated | + | |
| - | - While the loadlock pumps down, start Elphy Quantum and log in | + | - Place the transferstick in front of the loadlock with the excentric hole //down// |
| - | - Drive the stage to the exchange position with the '' | + | - Push the red button. The loadlock will be evacuated |
| - | - While the loadlock pumps down and the stage drives to the exchange position, turn the WD (Working Distance) to the '' | + | - While the loadlock pumps down, start Elphy Quantum and log in |
| - | | + | - Drive the stage to the exchange position with the '' |
| - | - Carefully slide the sampleholder onto the stage. **DO NOT USE FORCE, this is a delicate, micrometer accurate stage!** | + | - While the loadlock pumps down and the stage drives to the exchange position, turn the WD (Working Distance) to the '' |
| - | - Unscrew the transferstick // | + | |
| - | - // | + | - Carefully slide the sampleholder onto the stage. **DO NOT USE FORCE, this is a delicate, micrometer accurate stage!** |
| - | - Push the red button to vent the loadlock //while holding the transferstick so that it doesn' | + | - Unscrew the transferstick // |
| - | - If the transfer did not go very smooth or the pressure went up, put vacuum grease (next to the transfer stick) onto the transfer stick and slide it back and forth (you are still wearing gloves, so your hands will not get dirty) | + | - // |
| - | - Put the transfer stick in the holder | + | - Push the red button to vent the loadlock //while holding the transferstick so that it doesn' |
| - | - Turn the WD to the desired value (be very careful not too set it too far up, what 'too far' is depends on the sample holder; ask help if unsure | + | - If the transfer did not go very smooth or the pressure went up (strangely this is indicated by having no illumination of the current meter and you can't turn on main power), put vacuum grease (next to the transfer stick) onto the transfer stick and slide it back and forth (you are still wearing gloves, so your hands will not get dirty) |
| - | - Drive the stage to the chessy | + | - Put the transfer stick in the holder |
| - | - Turn the monitor contrast and brightness to 2 o' clock | + | |
| - | - PCD '' | + | |
| - | - Mode '' | + | - Drive the stage to the chessy |
| - | - Blanker '' | + | - Turn the monitor contrast and brightness to 2 o' clock |
| - | - SEI detector '' | + | - PCD '' |
| - | - Scan speed '' | + | - Mode '' |
| - | - Check that filament heating current setpoint = 0 | + | - Blanker '' |
| - | - Check that HV setpoint is 20 kV (or lower) | + | - SEI detector '' |
| - | - Switch on HV (red switch) | + | - Scan speed '' |
| - | - Switch current meter to 1 nA full scale | + | - Check that filament heating current setpoint = 0 |
| - | - Slowly turn filament heating current to the indicated value (at the ' | + | - Check that HV setpoint is 20 kV (or lower) |
| - | - You should now measure a current on the sample | + | - Switch on HV (red switch) |
| - | - Adjust contrast and brightness | + | - Switch current meter to 1 nA full scale |
| - | - You should now see the chessy | + | - Slowly turn filament heating current to the indicated value (at the ' |
| + | - You should now measure a current on the sample | ||
| + | - Adjust contrast and brightness | ||
| + | - You should now see the chessy | ||
| Line 59: | Line 64: | ||
| You can navigate with the joystick and set 1 of the 4 moving rates. (Be careful not to push '' | You can navigate with the joystick and set 1 of the 4 moving rates. (Be careful not to push '' | ||
| - | == Imaging == | + | == Imaging |
| Correction of astigmatism is most conveniently done on the chessy. Then you can move to the sample, if it is as thick as the chessy there should hardly be any astigmatism on the sample and you only need to adjust the focus a little bit. Taking a picture can be done in Elphy Quantum by clicking '' | Correction of astigmatism is most conveniently done on the chessy. Then you can move to the sample, if it is as thick as the chessy there should hardly be any astigmatism on the sample and you only need to adjust the focus a little bit. Taking a picture can be done in Elphy Quantum by clicking '' | ||
| Line 70: | Line 75: | ||
| To correct astigmatism turn the focus knob back and forth, if you see horizontal and vertical (or: these rotated over some angle) lines get out of focus at different points you have to adjust the stigmator. Set focus in the middle between where the horizontal and vertical features are sharp. Optimize X and Y astigmatism. Optimize focus. Repeat until no further improvement is seen. | To correct astigmatism turn the focus knob back and forth, if you see horizontal and vertical (or: these rotated over some angle) lines get out of focus at different points you have to adjust the stigmator. Set focus in the middle between where the horizontal and vertical features are sharp. Optimize X and Y astigmatism. Optimize focus. Repeat until no further improvement is seen. | ||
| - | Changing the aperture.\\ | + | == Changing the aperture |
| For currents below 100 pA, use the smallest aperture; between 100 pA and 5 nA use the second smallest and above 5 nA (hardly used) use the third smallest aperture. After changing the aperture, check the alignment by pushing the wobbler button and adjusting the X and Y position of the aperture until the center of the image remains fixed (also at large magnification and max amplitude of the wobbler). Then switch off the wobbler. | For currents below 100 pA, use the smallest aperture; between 100 pA and 5 nA use the second smallest and above 5 nA (hardly used) use the third smallest aperture. After changing the aperture, check the alignment by pushing the wobbler button and adjusting the X and Y position of the aperture until the center of the image remains fixed (also at large magnification and max amplitude of the wobbler). Then switch off the wobbler. | ||
| - | === Shut down === | + | === Shutting |
| - PCD on (button in) | - PCD on (button in) | ||
| - Turn back the filament heating current //slowly// to avoid breaking it | - Turn back the filament heating current //slowly// to avoid breaking it | ||
jeol_820_sem.1208359000.txt.gz · Last modified: 2008/04/16 15:16 by verstoep