 ===== Quantum Design service notes ===== ===== Quantum Design service notes =====
  
-* Connector test +* ** Connector test** \\ 
-{{:equiment:squid:mpms_black_lemo.pdf|}}+general test \\ 
 +{{:equiment:squid:mpms_black_lemo.pdf|}} \\
 {{:equiment:squid:mpms_probe_resistances_all.pdf|}} {{:equiment:squid:mpms_probe_resistances_all.pdf|}}
  
-* Flow test+* **Flow test** \\ 
 +check if cooling heating isn't fast (10K/min) \\
 {{:equiment:squid:high-power-impedance-test.mv.pdf|}} {{:equiment:squid:high-power-impedance-test.mv.pdf|}}
 +
 +* **Model 1802 heater/valve driver output** \\
 +check if cooling heating isn't fast (10K/min) \\
 +{{:equiment:squid:1014-307_output_drivers_m1802_multivu.pdf|}}
 +
 +* **Model 1802 fuses** \\
 +check if cooling heating isn't fast (10K/min) \\
 +{{:equiment:squid:1802-which_board_is_where_1.jpg‎?width=120px}}
 +{{:equiment:squid:m1802_rare_panel_fuses.jpg?width=120px}}
 +{{:equiment:squid:power_board_fuses.jpg?width=120px}}
 +{{:equiment:squid:voltages-1802-power-supply.jpg?width=120px}}
 +{{:equiment:squid:voltages-1802-zoom.jpg?width=120px}}
 +
 +* **Kepco Powersupply Tests** \\
 +With the following procedutre one can test the kepco PS independantly from the MPMS electronics this way we find out if the is at the electronics or at the Kepco:
 +
 +Attached bench test should help to coarsely evaluate the Kepco's ability to deliver current. Disconnect the wires from MPMS Gas/Magnet Tray to Kepco rare panel, 
 +connect an external power supply and wires electrically as shown in the schematic. Ramping the external power supply from 0 to 5 volts should cause adequate 
 +output of the Kepco voltmeter and 0-50 ampers output of the current.
 +
 +Make sure the shunt will be at least 0.1 Ohm - otherwise the test will fail and the Kepco will rail through to max current output, immediately.
 +
 +---------------
 +
 +The following info is Only for experienced users who has no possibility to send it back.
 +
 +We usually see Kepco failures with the big 70 Amp, rectifying diodes, big capacitors or the 12 2N3055 transistors.
 +Your MPMS hardware manual binder should contain complete Kepco manuals that hold functional description, testing procedures, schematics and part lists for repair reference.
 +