This is an old revision of the document!
Table of Contents
Description
The Leybold Z-400 is an RF diode sputtering system equipped with 3 3“ targets. It has a background pressure of 1e-6 mbar and a deposition pressure of ca. 3e-2 - 8e-2 mbar. The 2 substrate tables are water cooled and can hold substrates up to 3”. The system is equipped with 3-channel gas blending for Ar, N2 and O2. H2S can also be mixed in. The turbo pumping system is automated. The substrate tables can be RF-biassed for RF bias sputtering or ion etching.
Manual
WARNING! there is only 1 way to damage the Z-400, that is by changing the HV target selector while HV is on. This will destroy the klystron tube and literally explode the capacitors surrounding it resulting in many months of downtime due to the very long delivery time of these expensive components.
Contact me if the target you need is not in the system.
Switch off the pumps
- Check that the mass flow controller valve is closed (indicated by Close in the software).
- Close the green valve while holding the tubing (until the arrow points in the middle of the red marker). It needs to be tight, the valve has a end stop.
- Verify that cooling water is switched off (blue shower button is not illuminated).
- Check that the HV is off (two on/off buttons) and the power supply is off (red turning knob).
- Switch off turbo and rotary pump (1 switch).
Vent the chamber
- Decelerate the turbo by slowly venting to 5 mbar using the needle valve and closing it at 5 mbar. WAIT 5 MINUTES.
- Now, vent the chamber completely to atmospheric pressure and close the vent valve.
- Open the flip-top lid.
Mount substrates
- The right substrate table is preferred, use the one at the left if needed.
- You can use silver paint for thermal contact (dry time: >15 minutes), remove sample with a thin Gilette™ razor when you're done and clean substrate table with acetone.
- SEM stubs with samples on them can be loaded on the right table if your sample is not too thick (1 mm max).
Pump down
- Close the flip-top lid.
- Open the rotary pump valve ('Vor-Vakuum' button).
- When the pressure drops below 1e-2 mbar, open the green valve (use both hands) to keep the gas line clean.
- Pump down to the desired pressure (but always <1e-5 mbar).
Sputter
- Note background pressure in logbook.
- Switch on cooling water (blue button).
- Manual Process
- Select target with HV target selector knob (left of the chamber).
- Switch on power supply (red turning knob). It takes 2 minutes for the RF generator to warm up.
- Check the timings and positions for the (pre) sputtering, note this in the logbook.
- Set the target to the pre sputter position.
- Set the Argon flow setpoint (using the software, usually around 30-50 sccm) and set the MFC valve to 'Normal'.
- Check that the desired pressure is reached (typically 5E-3 mbar), if not adjust the flow accordingly.
- Set the DC voltage meter to 1 kV full scale (right display).
- Turn the HV setpoint to zero (black knob on power supply).
- Turn HV on (two on/off buttons).
- Set 500 V RF (2nd display from the left, 5 kV scale).
- Increase the flow setpoint temporarily to 75 sscm and check that the pressure rises.
- Wait until the plasma ignites (right display shows non-zero value).
- Quickly check that the ignited target is at pre sputter position (look through window, purple glow at front position).
- Next set the flow to the original value and check that the pressure stabilises at 5E-3 mbar.
- Set the DC potential to the desired value (typically 1 kV).
- With the pressure and the plasma stable, start your process (pre sputtering, sputtering with the correct timings and positions).
- When the process is finished (ignited target again at pre sputter position) set the HV setpoint to zero.
- Switch off HV (red
HV offbutton). - If you want to sputter a different material now, redo the procedure from step I.
- Automated Process (currently not implemented)
- …
- …
- …
End
- Close the MFC ('Close' button in software, read dialog, press OK).
- Switch off HV power supply (left red turning knob).
- Switch off cooling water.
- Check that the mass flow controller valve is closed (indicated in the software by 'Vlv: C').
- Close green valve (use both hands).
- Let the cooled substrates warm up before venting the chamber with N2 (see start of the manual).
Target Materials
The following materials are available:
|
|
Deposition rates Z406
| Material | Date | User | Pressure | DC Potential | Flow | Time | Measurement | Result | Rate | Notes |
|---|---|---|---|---|---|---|---|---|---|---|
| Al | 20120522 | Boltje | 1kV | 50 sccm Ar | 148 s | X-ray | 22.5 nm | 9.1 nm/min | ||
| Al | 20080514 | Jorina | 1kV | 25% Ar | 10min | X-ray | 80 nm | 8 nm/min | ||
| Al | 20100520 | Boltje | 4e-3 mbar | 500V | ~25% Ar | 10 min | X-ray | 25.6 nm | 2.6 nm/min | |
| Al | 20110606 | Boltje | 1kV | 50 sccm Ar | 4 min | X-ray | 38 nm | 9.5 nm/min | ||
| Al | 20151007 | Timothy | 4.9e-3 mbar | 1kV | 49 sccm | 3 min | X-Ray | 30.0 nm | 10.0 nm/min | |
| Al2O3 | 20120606 | Boltje | 1kV | 60 sccm Ar, 10% O2 | 5 min | X-ray | 9.2 nm | 1.8 nm/min | ||
| Al2O3 | 20050127 | Gertjan | 1kV | 25% Ar | 2min | RBS | 8 nm | 4 nm/min | ||
| Al2O3 | 20100519 | Boltje | 4e-3 mbar | 1kV | ~25% Ar | 15 min | X-ray | 32.4 nm | 2.2 nm/min | |
| Al2O3 | 20111028 | Boltje | 0.5 kV | 50 sccm Ar | 15min | X-ray | 10.7 nm | 0.71 nm/min | ||
| Ag | 20111014 | Boltje | 1kV | 50 sccm Ar | 2min | X-ray | 38.2 nm | 19.1 nm/min | ||
| Au | 20120404 | Boltje | 1kV | 50 sccm Ar | 3min | X-ray | 45.6 nm | 15.2 nm/min | ||
| Au | 20100330 | Boltje | 1kV | ~25 Ar | 3min | X-ray | 54.3 nm | 18.1 nm/min | ||
| Au | 20110202 | Boltje | 1kV | 50 sccm Ar | 2min | X-ray | 29.1 nm | 14.6 nm/min | ||
| Au | 20111206 | Boltje | 1kV | 50 sccm Ar | Xmin | X-ray | 51.4 nm | 15.0 nm/min | ||
| Au | 20120402 | Boltje | 1kV | 50 sccm Ar | 80sec | X-ray | 19.2 nm | 14.4 nm/min | ||
| Co | 20111207 | Boltje | 1kV | 50 sccm Ar | 545sec | X-ray | 63.1 nm | 6.95 nm/min | ||
| Co | 20100331 | Boltje | 1kV | 25% Ar | 3min | X-ray | 16.5 nm | 5.5 nm/min | ||
| Co | 20110104 | Boltje | 1kV | 50 sccm Ar | 10min | X-ray | 63.9 nm | 6.4 nm/min | ||
| Cr | 20090709 | Boltje | 1kV | 25% Ar | 4min | X-ray | 30.0 nm | 7.5 nm/min | ||
| Cr | 20141111 | Boltje | 3e-5 mbar | 1kV | 48 sccm | 80 s | X-Ray | 12.8 nm | 9.6 nm/min | |
| Cu | 20120606 | Boltje | 1kV | 50 sccm Ar | 3min | X-ray | 37.6 nm | 12.5 nm/min | ||
| Cu | 20100203 | Boltje | 1kV | 25% Ar | 5min | X-ray | 63 nm | 12.6 nm/min | ||
| Cu | 20110111 | Boltje | 1kV | 50 sccm Ar | 5min | X-ray | 57.6 nm | 11.5 nm/min | ||
| Cu | 20111206 | Boltje | 1kV | 50 sccm Ar | 5min | X-ray | 58.2 nm | 13.4 nm/min | ||
| Cu | 20120404 | Boltje | 1kV | 50 sccm Ar | 5min | X-ray | 51.5 nm | 10.3 nm/min | ||
| Cu-etch | 20120606 | Boltje | 50 sccm Ar | 3 min | X-ray | 2.1 nm | 0.7 nm/min | 60W | ||
| Cu99.5Bi0.5 | 20150429 | Boltje | 3e-5 mbar | 1kV | 50 sccm Ar | 120s | X-Ray | 22.0 nm | 11.0 nm/min | |
| Co | 20120404 | Boltje | 1kV | 50 sccm Ar | 10min | X-ray | 68.0 nm | 6.8 nm/min | ||
| Fe | 20100407 | Boltje | 1kV | 25% Ar | 4min | X-ray | 22.0 nm | 5.5 nm/min | ||
| MoGe | 2007 | Edoardo | 1kV | 25% Ar | 15mins* | X-ray | 84.2 nm | 5.6 nm/min | * no pauses, silver paint | |
| MoGe | 20080204 | Ivan & Edo | 1kV | 25% Ar | 18 mins* | X-ray | 86 nm | 4.8 nm/min | *pauses (1min) | |
| MoGe | 20110415 | Boltje | 1kV | 50 sccm Ar | 10min | X-ray | 55.3 nm | 5.5 nm/min | ||
| MoGe* | 20110816 | Boltje | 1kV | 50 sccm Ar | 10min | X-ray | 51.5 nm | 5.2 nm/min | ||
| MoGe | 20170327 | Jean-Pierre | 4.9e-3 mbar | 1kV | 49 sccm | 25 min | Profilometer | 137 nm | 5.5 nm/min | |
| Nb65Ge35 | 2005 | Bas v Leewuwen | 1kV | 25% Ar | - | X-ray | - | 4.89 nm/min | ||
| Nb65Ge35 | 2004 | Tibi | <3e-6 mbar | 1kV | 25% Ar | 6 min | X-ray | 28 nm | 4.67 nm/min | |
| NbN | 20111212 | Boltje | 1.0 kV | 18% Ar, 10% N2* | x min | X-ray | 71.3 nm | 3.56 nm/min | * no blend | |
| NbN | 20050126 | Gertjan | 1.7 kV | 18% Ar, 10% N2* | 8 min | RBS/X-ray | 60/75nm | 7.5-9 nm/min | * blend | |
| NbN | 20050124 | Olaf | 1.0 kV | 18% Ar, 10% N2* | 1356s | X-ray | 49.5 nm | 2.2 nm/min | * no blend | |
| NbN | 20080310 | Jorina | 1.0 kV | 18% Ar, 10% N2* | 22min 42s | X-ray | 55.15 nm | 2.4 nm/min | * no blend | |
| Ni | 20110114 | Boltje | 1kV | 50 sccm Ar | 9min | X-ray | 58.8 nm | 6.5 nm/min | ||
| Ni | 20140325 | Boltje | 3e-5 mbar | 1 kV | xx sccm | 5 min | X-Ray | 43.5 nm | 8.70 nm/min | |
| NiGd | 20051024 | ? | 1kV | 25% Ar | X-ray | 67.5 nm | 6.8 nm/min | |||
| Pt | 20110307 | Boltje | 0.5kV | 50 sccm Ar | 10 min | X-ray | 15.6 nm | 1.56 nm/min | ||
| Pt | 20120612 | Boltje | 1kV | 50 sccm Ar | 10 min | X-ray | 67.9 nm | 6.8 nm/min | ||
| Pt | 20061113 | Chris | 0.5kV | 25% Ar | 2mins | X-ray | 5.0 nm | 2.5 nm/min | ||
| Py | 20161223 | Mechielsen | 1kV | 50 sccm Ar | AFM | 7.9 nm/min | ||||
| Py | 20170220 | Casper | 4.8e-3 mbar | 1kV | 49 sccm | xx min | X-ray | 585 nm | 7.8 nm/min | |
| Si | 20150422 | Boltje | 5e-3 mbar | 1 kV | 49 sccm | 3 min | X-Ray | 23.0 nm | 7.66 nm/min | |
| Si3N4 | 20120522 | Boltje | 1kV | 50 sccm Ar | 469 s | X-ray | 36.0 nm | 4.6 nm/min | ||
| Si3N4 | 20110609 | Boltje | 1kV | 50 sccm Ar | 8 min | X-ray | 31.8 nm | 4.0 nm/min | ||
| Si3N4 | 20120522 | Boltje | 1kV | 50 sccm Ar | 469 s | X-ray | 36.0 nm | 4.6 nm/min | ||
| SiO2 | 20131114 | Boltje | 5e-3 mbar | 1kV | 38 sccm | 5 min | X-Ray | 15.4 nm | 3.08 nm/min | |
| Ti | 20160428 | Annette | 5.1e-3 mbar | 1kV | 49 sccm | 10 min | profilometer | 30 nm | 3.0 nm/min | |
| W | 20060116 | Vincent | 1kV | 25% Ar | 10 min | X-ray | 38.5 nm | 3.9 nm/min |
* repaired target, with a few new Ge pieces
Deposition rates Z407
| Material | Date | User | Pressure | DC Potential | Flow | Time | Measurement | Result | Rate | Notes |
|---|---|---|---|---|---|---|---|---|---|---|
Recipes & Instructions
Miscellaneous
The digital version of the log sheet: z400logbook.pdf